Page 190 Troubleshooting 2008T Machine Operator’s Manual P/N 490122 Rev N Emptying too long The dialyzer empty program has Alarm If blood is not sensed, return the dialyzer supply and return lines to exceeded its maximum limit. The shunt and close the shunt door to terminate the program. If the machine was in dialysis (blood sensed), turn machine power off and back on to clear the program. Fresenius 2008 Machine Manual The 2008K hemodialysis machine is manufactured by: Fresenius USA, Inc. 4040 Nelson Avenue Concord, CA 94520 Installation, maintenance, calibration and other technical information may be. Finally, the manual leads the reader through the two communication protocols that may be used with a 2008 series hemodialysis machine: standard and checksum. Also included in the appendices are details about the data that may be sent to the machine and the expected information returned by the machine. Page 1 2008T Hemodialysis Machine Operator’s Manual Caution: Federal (US) law restricts this device to sale only by or on the order of a physician P/N 490122 Rev N.; Page 2 Contact Fresenius Medical Care Technical Support for applicable Field Service Bulletins. The spare parts manual for the model 2008T and other information may be found on our website at www.fmcna.com Caution: Federal (US. Read PDF Fresenius 2008 Machine Manual sale by or on the order of a physician. As part of the Conditions for Coverage, it is the medical director’s responsibility that the operators of the Fresenius Medical Care 2008® Series hemodialysis machines, bibag and 2008 Series Spare Parts Manual - Fresenius Medical Care.

Fresenius 2008t dialysis machine manualFresenius 2008 dialysis machine manual transmission

Fresenius 2008t Dialysis Machine Manual

I'm having a problem with a 2008K machine in regards to 'high' TMP (technically low because TMP is a negative measurement). It does not go high enough to alarm, but it runs around 360-420 during a normal treatment and while in the technical office during troubleshooting. This is without the UF pump turned on. I have verified all calibrations numerous times. I have tried swapping in a whole card cage including cables to rule out electronics, which resulted in the problem following the hydraulics, so I know it's not any of the boards (except maybe the distribution board). I have tried swapping the entire balancing chamber (including valves)and hydroblock with known good ones. I have tried bypassing the heat exchanger, tried a new pressure transducer, verified the pump calibrations/incoming water pressure/dialysate pressure/venous pressure/flow, dear, and loading pressures/ all numerous times, even recently and all are good. I have replaced the CFS switch and the entire air separator assembly (valve 43 included) with new parts to rule those out. At this point I'm at a complete loss.
As a side note, I have tried using the troubleshooting guide. Unless you have a problem that causes an alarm state (which I don't, just TMP running higher than it normally should) the guide is next to useless.
I've started heading in the direction of a restriction from a valve starting with bypassing valve 25. I'm looking for some ideas here. Help!!! ;-)